Calibrate dew point analyzers and transmitters in-house, against a chilled mirror reference, instead of shipping them away.

Who needs it

  • Instrument service laboratories that calibrate customers' moisture analyzers.
  • Plants with many dew point points — gas processing, air separation, compressed air — that want short calibration turnaround.
  • Metrology institutes and quality departments that need a traceable reference for sub-zero dew points.
  • Distributors that want to offer calibration alongside instrument sales.

In many markets few commercial laboratories calibrate below 0 °C dew point, so analyzers are sent abroad and are out of service for weeks. An in-house system removes that wait.

System

YKDC220 low-humidity calibration system

A 19-inch rack system with a separately installed compressor. It combines a dry gas generator, a low-humidity generator and a chilled mirror reference hygrometer with the connecting gas lines.

Ask for a configuration and quote

How the system works

The DC1000 makes very dry air; the DC2000 mixes dry and wet gas to the set dew point; the reference hygrometer and the instruments under test see the same gas.

Compressor systemYK-2×750-55Oil-free, pre-dryer Dry gas generatorDC1000 · PSA≤ 1 ppmV (−75 °C DP) Low-humidity generatorDC2000 · MFC mixingSet point −20.0 °C FP Reference hygrometerDM9700 / DM9800 / DM9900Reading −20.0 °C stable Instrument under testAnalyzer or transmitterReading −19.4 °C Δ +0.6 Illustration: set points step −20 → −40 → −60 °C frost point; readings are simulated.

On a phone, swipe the diagram sideways.

ComponentRoleKey data
YK-2×750-55Compressed air supplyFully enclosed, oil-free compressor with pre-dryer and sound-muffling cabinet
DC1000Dry gas generatorPressure swing adsorption; up to 10 NL/min of dry air at 1 ppmV or less (−75 °C dew point); 1/4" stainless steel compression fittings
DC2000Low-humidity generatorFlow-divided mixing with four electronic mass flow controllers; fluctuation less than ±0.2 °C frost point; multiple user-configurable calibration points; three outlets (one reference, two calibration)
DM9700 / DM9800 / DM9900Reference hygrometerChilled mirror; DM9900 to −90 °C frost point, better than ±0.2 °C near −90 °C; calibration certificate available (chargeable)

Typical procedure

  1. Purge the lines with dry gas from the DC1000 until the reference reads stable.
  2. Set the first calibration point on the DC2000; wait for the reference and the instruments under test to stabilise.
  3. Record reference and instrument readings; repeat at each user-configured point.
  4. Adjust or certify each instrument against the reference readings.

The system is designed to the equipment set required by the Chinese verification regulation JJG 499-2004 for precision dew point meters. Your quality system defines the points and acceptance limits.

Scope
  • The YKDC220 is configured and quoted per project, including the choice of reference hygrometer.
  • The calibrated range depends on the reference chosen: DM9700 to −70 °C, DM9800 to −80 °C, DM9900 to −90 °C frost point.