Brewery CO₂ recovery: from fermenter to filling hall

Follow fermentation CO₂ through scrubbing, compression, drying and liquefaction back to carbonation. Tap each point to see what is measured and which YORK analyzer fits.

Breweries in Denmark, Vietnam, Thailand and elsewhere recover the CO₂ from fermentation, purify it and use it again for carbonation, instead of buying it in. Recovery pays back only if the gas is collected at the right time and dried and purified to beverage grade.

FERMENTATION GASFERMENTERSFOAM TRAP /SCRUBBERCOMPRESSORCARBONDRYERLIQUEFIERLIQUID CO₂VAPORIZERCARBONATION /FILLING HALL1CO₂ %2H₂O3H₂O4O₂ ppm5CO₂ ppm
YORK instrumentIntegrated system (supplied through YORK)

Tap a numbered point. Swipe the diagram sideways on a phone.

Request a quote for this point →
Auto tour
Project list 0 →

How the process works

  1. Fermentation. Yeast produces CO₂; early gas still contains air.
  2. Scrubbing. A foam trap and water scrubber remove foam and water-soluble impurities.
  3. Compression and purification. The gas is compressed, passed through activated carbon and dried.
  4. Liquefaction. Dry CO₂ is liquefied and stored.
  5. Reuse. A vaporizer supplies CO₂ to carbonation, filling and purging.
About the values

Values are typical figures from published specifications and industry practice. Your plant, licensor or grid specification always governs.

Measurement points at a glance

#PointMeasuresTypical value / limitYORK instrument
1Fermenter off-gasCO₂ concentrationRecovery starts once the off-gas is rich enough in CO₂; the threshold is set by the recovery plant supplierPGA520
2After the dryer, before the liquefierMoistureVery dry to avoid ice in the liquefierDPT830
3CO₂ to carbonationMoisture in the CO₂Beverage-grade specifications commonly limit water to about 20 ppm by volume (about −55 °C frost point at 1 atm)DPT-600 Plus
4Recovered CO₂ purityTrace oxygen and other impuritiesBeverage-grade specifications set oxygen at the ppm level; integrated system: trace oxygen analyzer for CO₂ service
5Filling hall and cellar airCO₂ in room air (safety)Workplace exposure limit 5,000 ppm (8-hour average) in the UK and many EU countriesPGA520

Questions we are asked

Can YORK verify beverage-grade CO₂ purity?

YORK instruments check moisture. Trace oxygen and other impurities at ppm level need dedicated analyzers, available as an integrated system.

Why not measure oxygen in CO₂ with an electrochemical cell?

Electrochemical cells are not used in a CO₂ background. Percent-level oxygen in CO₂ needs a paramagnetic analyzer; ppm levels need a trace analyzer.

What CO₂ level is unsafe in a cellar?

The UK and many EU countries set a workplace exposure limit of 5,000 ppm averaged over 8 hours. Alarm levels follow your safety assessment.

Discuss your application